Skip to content

Public FDA record

Z-0907-04: Surgical Tubing Pack, Model: AMO Prestige Pack, Product Code OPO40

Recall record Z-0907-04 for Allergan Medical Optics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0907-04
Company in FDA record
Allergan Medical Optics Inc
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The lots are being recalled because the tray lid may not be adequately sealed under the red arrow in the upper right corner of the package tray. The blue wrapping within the tray may not be sterile and may pose a potential risk of infection to the patient if used in a sterile field.

Source facts

Action
On March 31, 2004, a product notification letter was mailed to all customers via certified mail. The customers were instructed to stop using, quarantine and return all units. They were instructed to complete the Facsimile form provided and faxed it to the AMO Customer Service at 888-324-0021. They were also instructed to call AMO Customer Service at 800-366-6554 to make arrangement to return the product.
Initiated
2004-03-29
Root cause
Other
Terminated
2004-11-04
Distribution
Distribution was to 67 domestic customers and internationally.
Recall number
Z-0907-04
Recall status
Terminated
CDRH record ID
32748
FDA classified
2004-07-20
Recall event number
28904

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗