Philips Medical Systems North America Co. Phillips
FDA classification date
Sep 25, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Units do not comply with the requirement to provide automatic beam limitation within +/- 3 degrees of horizontal, vertical, and perpendicular to the plane of the image receptor during radiographic procedures.
Source facts
Action
On 7/23/07 Philips mailed a letter dated July 9, 2007 titled Electronic Product Radiation Warning. The letter advises customers of the issue and asks that customers apply the enclosed cautionary label to the device. The letter also explains that there will be a system modification to be installed. On 8/20/07 the firm sent the same letter, along with a letter dated 8/20/07 to customers thought to have either scrapped their units and the customer had previously indicated they were no longer using the unit. The 8/20/07 letter requested the customer contact Philps if they are currently using the device.
Initiated
2007-07-23
Root cause
Other
Terminated
2012-09-17
Distribution
Units were distributed to 1543 hospitals and medical centers throughout the US. No international distribution from this firm.
Recall number
Z-0907-2007
Recall status
Terminated
CDRH record ID
64513
FDA classified
2007-09-25
Recall event number
44910