Skip to content

Public FDA record

Z-0907-2010: Aquilion One Whole Body X-Ray CT Scanner Whole body computed tomography scanning

Recall record Z-0907-2010 for Toshiba American Medical Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0907-2010
Company in FDA record
Toshiba American Medical Systems Inc
FDA classification date
Mar 2, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

When W-Volume is performed in Prospective CTA or Target CTA studies with ECG gating, unnecessary intervals may be set before starting the volume scan. As a result, scanning cannot be performed at the appropriate timing of constant enhancement and clinically useful images may not be obtained.

Source facts

Action
A corrective measure will be installed to prevent measures to avoid this phenomenon until corrective actions have been taken. Customers are advised: (1) When performing scanning in Prospective CTA or Target CTA studies with ECG gating, do not use W-Volume scanning but use Helical Scanning. (2) Please share this information with all users and reviewing radiologists as well as clinical engineering or Biomedical group at facility (3) Please complete and return the attached form and fax it to the toll-free number at the top or by email to raffairs@tams.com. Contact: Paul Biggins
Initiated
2010-01-07
Root cause
Device Design
Terminated
2012-09-12
Distribution
Nationwide: AL AR AZ CA CO FL GA IA MA MD ND NJ NV NY OH PA PR andVA
Recall number
Z-0907-2010
Recall status
Terminated
CDRH record ID
87967
FDA classified
2010-03-02
Recall event number
54258

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗