Open original FDA record ↗Recall record
- FDA identifier
- Z-0907-2010
- Company in FDA record
- Toshiba American Medical Systems Inc
- FDA classification date
- Mar 2, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
When W-Volume is performed in Prospective CTA or Target CTA studies with ECG gating, unnecessary intervals may be set before starting the volume scan. As a result, scanning cannot be performed at the appropriate timing of constant enhancement and clinically useful images may not be obtained.
Source facts
- Action
- A corrective measure will be installed to prevent measures to avoid this phenomenon until corrective actions have been taken. Customers are advised: (1) When performing scanning in Prospective CTA or Target CTA studies with ECG gating, do not use W-Volume scanning but use Helical Scanning. (2) Please share this information with all users and reviewing radiologists as well as clinical engineering or Biomedical group at facility (3) Please complete and return the attached form and fax it to the toll-free number at the top or by email to raffairs@tams.com. Contact: Paul Biggins
- Initiated
- 2010-01-07
- Root cause
- Device Design
- Terminated
- 2012-09-12
- Distribution
- Nationwide: AL AR AZ CA CO FL GA IA MA MD ND NJ NV NY OH PA PR andVA
- Recall number
- Z-0907-2010
- Recall status
- Terminated
- CDRH record ID
- 87967
- FDA classified
- 2010-03-02
- Recall event number
- 54258