Siemens became aware of a potential safety issue when using the "In-session resumption" feature of the syngo RT Therapist (RTT) 4.3 software. The RTT will send again the previously completed beam or segment for delivery to the control console, thus repeating the already delivered Monitor Units of the same beam or segment.
Source facts
Action
The firm issued "Urgent Medical Device Correction" letters via mail to all its direct consignees, informing them of the affected product issue and providing instructions on the recall. The firm stated that the release of this UI TH005/13/S to distribute the Customer Safety Advisory Notice began on January 17, 2013.
Initiated
2013-01-17
Root cause
Software design
Terminated
2013-03-05
Distribution
5 devices, all located outside of the United States in Germany.
Recall number
Z-0907-2013
Recall status
Terminated
CDRH record ID
115590
FDA classified
2013-03-05
Recall event number
64214