Z-0907-2022: Philips Respironics V60 Ventilator (All Models, including: 1053613, 1053614, 1053615, 1053616, 1053617, 1

Recall record Z-0907-2022 for Respironics California, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0907-2022
Company in FDA record
Respironics California, LLC
FDA classification date
Apr 18, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Ventilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.

Source facts

Action
On March 14, 2022, Respironics California, LLC issued an Urgent Medical Device Correction notification to all affected consignees. On April 28, 2022 Respironics issued a revised customer communication titled Update: Urgent Medical Device Correction, dated April 20, 2022, which described mitigating actions. On June 30, 2022, Respironics issued an Updated Urgent Medical Device Correction Notification. This letter was to provide customers with an update on Philips Respironics actions to address an issue related to the internal source (35V Rail) that powers the V60/V60 Plus ventilators. In addition, between the dates of January 23, 2023 - March 8, 2023, Respironics issued a "Field Safety Notice" to customers who own the V680 model. the notification informed customers that the device will be removed from the market. The communication contained the following listed below: To address the issue with the 35V Rail, Philips Respironics will be deploying a technical solution (changing two resistors within the ventilator) that will cause the V60/V60 Plus ventilators to alarm in all cases should it experience an issue with the 35V Rail. Philips Respironics will be contacting customers to schedule an appointment to implement the technical solution in your V60/V60 Plus ventilator(s) starting in October 2022. Philips Respironics is offering customers a loaner EV300 ventilator while the customer awaits deployment of the technical solution in its V60/V60 Plus ventilator(s) should it be unable to implement one of the mitigations provided in the April 2022 letter. All other information provided in the April 2022 customer communication is unchanged. This notice needs to be provided to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred.
Initiated
2022-02-28
Root cause
Device Design
Distribution
Worldwide distribution US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of Afghanistan, Albania, Algeria, American Samoa, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Comoros, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guam, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jersey, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland , Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, St. Pier, Miquel., Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Kingdom, Utd. Arab. Emir., Venezuela, Vietnam, Wallis, Futuna, Zambia, Japan.
Recall number
Z-0907-2022
Recall status
Open, Classified
CDRH record ID
192390
FDA classified
2022-04-18
Recall event number
89767

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗