Open original FDA record ↗Recall record
- FDA identifier
- Z-0907-2025
- Company in FDA record
- Hologic, Inc.
- FDA classification date
- Jan 14, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential weak pouch seal on Panther Fusion GBS assay cartridge, causing potential for invalid or incorrect results leading to a possible delay of treatment.
Source facts
- Action
- An Urgent Medical Device Recall Notification letter issued 12/13/2024 via Fed Ex. The letter instructs customers to discontinue use of the Panther Fusion GBS Assay kit lot 715927 containing cartridge lot 624907 and segregate and count the number of affected Panther Fusion GBS Assay kits from lot 715927 containing cartridge lot 624907. Customers are instructed to destroy the product and fill out the online Customer Response Form (CRF) indicating that the on-hand inventory of recalled Panther Fusion GBS Assay Kit has been destroyed. If you have any questions or concerns regarding this recall notification, please contact Hologic Technical Support at +1 888 484 4747 or +1 858 410 8511 during the hours of 5 am PT and 5 pm PT, or by e-mail at molecularsupport@hologic.com.
- Initiated
- 2024-12-13
- Root cause
- Process control
- Distribution
- AR, CA, CT, FL, IN, KS, LA, MA, ME, MI, NC, NJ, NV, NY, PR, TX, UT, VT, WA, WV, and Canada.
- Recall number
- Z-0907-2025
- Recall status
- Open, Classified
- CDRH record ID
- 211613
- FDA classified
- 2025-01-14
- Recall event number
- 95948