Skip to content

Public FDA record

Z-0908-2010: Rubella IgG EIA test kit, containing 1 Rubella IgG Microplate, 1 Negative Control, 1 High Positive Contro

Recall record Z-0908-2010 for Bio-Rad Laboratories. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0908-2010
Company in FDA record
Bio-Rad Laboratories
FDA classification date
Mar 2, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Elevated optical density (OD) values with the kit Negative Control.

Source facts

Action
An "Urgent Product Correction" letter dated August 26, 2008 was sent via overnight carrier to all customers. The letter described the affected product, problem, and action to be taken by customer. The customers are to review their current inventory of the product for any remaining kits of the specified lot number. The customer is to complete the form and fax to the Technical Support department if replacement kits were needed. The remaining kits will be replaced once the form is completed and faxed. If you have any questions, please contact Technical Support at 1-800-224-6723, options 2 then 4.
Initiated
2008-08-20
Root cause
Employee error
Terminated
2010-04-08
Distribution
Nationwide distribution: NE, NY, FL, GA, MA, OR, WI, MN, MI, MS, NC, CA, MD, MO, IA, DC
Recall number
Z-0908-2010
Recall status
Terminated
CDRH record ID
73408
FDA classified
2010-03-02
Recall event number
49449

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗