Z-0908-2014: Product Name: Model 7305- Mosaic Aortic Obturator
The Mosaic Aortic Obturators are used with the Mosaic
Recall record Z-0908-2014 for Medtronic Cardiovascular Surgery-the Heart Valve Division. View source facts, labeled dates, and the original FDA record.
Medtronic, Inc. initiated a recall of Medtronic Mosaic Aortic Obturator Model 7305
used with the Mosaic Porcine Aortic Bioprosthesis, Model Number 305,
because of higher-than-expected transvalvular gradients occurring post implant.
Source facts
Action
Medtronic sent an Customer Notification letter dated January 9, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers.
Medtronic recommends surgeons continue to remain diligent in selecting the appropriate size of Mosaic Aortic Bioprosthesis. Adoption of the updated Mosaic Aortic Bioprosthesis sizing chart (iEOA chart) and Mosaic Obturators/Sizers will facilitate optimal aortic valve size selection.
Your Medtronic Field Representative will be contacting you about receiving a new Mosaic Aortic Obturator/Sizer set and sizing chart. If you would prefer to order the new Mosaic Aortic Obturator/Sizer set directly through Medtronic Customer Service, contact them at 1-800-848-9300.
Medtronic will notify all applicable regulatory agencies, as required, about this matter. Please share this notification with others in your organization as appropriate. We appreciate your review of this notification and apologize for the inconvenience that it may cause. If you have any questions, please contact your Medtronic Field Representative or Lifeline Technical Services at 1-877-526-7890.
Initiated
2014-01-09
Root cause
Device Design
Terminated
2014-05-30
Distribution
Worldwide Distribution - USA (nationwide)
Recall number
Z-0908-2014
Recall status
Terminated
CDRH record ID
125069
FDA classified
2014-02-03
Recall event number
67242
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.