The power manager board may cause a short circuit, resulting in loss of system power, loss of central control monitor or electrical arcing.
Source facts
Action
An "Urgent Medical Device Correction" letter was sent to consignees on November 7, 2008. The letter described the reason for correction and provided instructions to customers. Terumo Cardiovascular Systems Corporation will replace the power manager boards on the affected devices. Customers were instructed to complete and fax the Customer Response Form to CVS Customer Service at 1-800-292-6551 (US only) or respond by telephone at 1-800-521-2818. Please direct questions to CVS Customers Service at 1-800-521-2818.
Initiated
2008-11-07
Root cause
Nonconforming Material/Component
Terminated
2009-07-07
Distribution
Worldwide Distribution including USA and countries of Australia, Belgium, Canada, Columbia, Costa Rica, Dominican Republic, Egypt, Germany, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arab Emirates and Viet Nam.
Recall number
Z-0909-2009
Recall status
Terminated
CDRH record ID
75749
FDA classified
2009-02-24
Recall event number
50554