Open original FDA record ↗Recall record
- FDA identifier
- Z-0909-2011
- Company in FDA record
- Accuray Inc
- FDA classification date
- Jan 18, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The product has the potential for the extension coupling of the Synchrony extension mounting hardware to become loose over time. The assembly weighs approximately 17 pounds; an unexpected contact with users or patients may result in serious injury if the boom descends unexpectedly.
Source facts
- Action
- Accuracy Inc., sent an URGENT ADVISORY NOTIFICATION letter, dated March 16, 2009, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inspect their system to see if it was affected. Customers were instructed to contact Accuracy Customer Support at 1-877-668-8667 (USA) or 1-408-716-4700 or e-mail customersupport@accuracy.com to schedule service.
- Initiated
- 2009-03-16
- Root cause
- Device Design
- Terminated
- 2011-01-19
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Turkey, Russia, France, Italy, the Netherlands, the UK, Spain, Greece, Switzerland, Germany, India, Taiwan, China, Hong Kong, Korea, Thailand, Malaysia and Japan
- Recall number
- Z-0909-2011
- Recall status
- Terminated
- CDRH record ID
- 80676
- FDA classified
- 2011-01-18
- Recall event number
- 51550