The device has a software anomaly which causes the operating system to 'lock up' and data to become corrupted during service procedures.
Source facts
Action
The recalling firm plans to notify their customers of this recall/soft ware correction during a scheduled visit to be conducted at each customer site by a Hitachi Field Service Technician. During the visit, the necessary corrective software will be installed. The visits/corrections began on: 2/21/06.
Initiated
2006-02-21
Root cause
Other
Terminated
2012-02-16
Distribution
Nationwide
Recall number
Z-0910-06
Recall status
Terminated
CDRH record ID
45241
FDA classified
2006-05-31
Recall event number
35087