Open original FDA record ↗Recall record
- FDA identifier
- Z-0910-2015
- Company in FDA record
- McKesson Israel Ltd.
- FDA classification date
- Jan 5, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
An ECG procedure time stamp is incorrect, due to a software error in the McKesson Cardiology ECG Management with software versions 13.1 and 13.1.1.
Source facts
- Action
- McKesson sent an Urgent Field Safety Notice dated December 16, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed how to identify the potentially affected procedures and update the procedure date (if relevant) until a software update is made advailable for installation by McKesson. McKesson will proactively identify prior procedures affected by this issue and will notify customers of any issues identified. In addition to the Field Safety Notice, McKesson Israel will provide correction to the affected SW versions and will contact customers for deployment of a software patch that prevents this scenario from occurring. Customers with questions should contact McKesson Support at 1-866-777-0202. For questions regarding this recall call 404-338-3556.
- Initiated
- 2014-12-19
- Root cause
- Software design
- Terminated
- 2016-02-25
- Distribution
- Worldwide Distribution - USA including MA, MS, NH, NC, TX, WA and Internationally to the United Kingdom.
- Recall number
- Z-0910-2015
- Recall status
- Terminated
- CDRH record ID
- 132318
- FDA classified
- 2015-01-05
- Recall event number
- 70144