Z-0910-2026: AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product Description: The AMSCO 7053HP Was

Recall record Z-0910-2026 for Steris Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0910-2026
Company in FDA record
Steris Corporation
FDA classification date
Dec 11, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).

Source facts

Action
On November 19, 2025, firm began notifying customers via Urgent Medical Device Field Correction letters. All affected customers will be contacted by a STERIS Service Technician to arrange an onsite inspection of their units to perform the correction. If the affected unit has not been installed yet, the correction will be performed as part of the installation process. Users can continue to use their units prior to completion of the correction. Under normal operation, the system is designed to detect such electrical arcing specifically, indicating a Drying Over Temperature Switch Tripped alarm. Should your unit alarm, abort the cycle, turn off power to the unit, and contact STERIS.
Initiated
2025-11-19
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution in the states of Arizona, Colorado, Florida, Georgia, Illinois, Kentucky, Louisiana, Michigan, New Hampshire, New York, North Carolina, Pennsylvania, Tennessee, Texas, Utah, Victoria, Virginia, Washington, Wisconsin and the countries of Australia.
Recall number
Z-0910-2026
Recall status
Open, Classified
CDRH record ID
217381
FDA classified
2025-12-11
Recall event number
97986

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗