Skip to content

Public FDA record

Z-0911-2026: DEXLOCK Achilles Repair Implant Kits, MAKT4520

Recall record Z-0911-2026 for Medline Industries, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0911-2026
Company in FDA record
Medline Industries, LP
FDA classification date
Dec 11, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There have been multiple complaints of drill bit fusing to the bushing/sleeve within the drill guide during use. Retrieving a replacement drill bit and drill guide may prolong the surgical procedure.

Source facts

Action
Firm began notifying consignees on 11/11/2025 via Urgent Medical Device Recall letters. Customers have been instructed to over-label kits with a sticker instructing them to not use the drill guide within the Achilles Kit and to use the reusable 3.6mm drill guide from the Soft Tissue Instrument Tray. Customers who do not have a Soft Tissue Instrument Tray have been instructed to notify Medline so that a tray can be provided. Customers were instructed to quarantine all affected product in stock and return a completed response form to Medline. Upon receipt of the response form, the firm will then provide over-labels to place on affected inventory. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Initiated
2025-11-11
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CO, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MT, NC, NJ, OH, OK, PA, SC, TX, VA, WI.
Recall number
Z-0911-2026
Recall status
Open, Classified
CDRH record ID
217513
FDA classified
2025-12-11
Recall event number
98056

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗