Open original FDA record ↗Recall record
- FDA identifier
- Z-0911-2026
- Company in FDA record
- Medline Industries, LP
- FDA classification date
- Dec 11, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There have been multiple complaints of drill bit fusing to the bushing/sleeve within the drill guide during use. Retrieving a replacement drill bit and drill guide may prolong the surgical procedure.
Source facts
- Action
- Firm began notifying consignees on 11/11/2025 via Urgent Medical Device Recall letters. Customers have been instructed to over-label kits with a sticker instructing them to not use the drill guide within the Achilles Kit and to use the reusable 3.6mm drill guide from the Soft Tissue Instrument Tray. Customers who do not have a Soft Tissue Instrument Tray have been instructed to notify Medline so that a tray can be provided. Customers were instructed to quarantine all affected product in stock and return a completed response form to Medline. Upon receipt of the response form, the firm will then provide over-labels to place on affected inventory. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
- Initiated
- 2025-11-11
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution in the states of AL, AR, AZ, CO, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MT, NC, NJ, OH, OK, PA, SC, TX, VA, WI.
- Recall number
- Z-0911-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217513
- FDA classified
- 2025-12-11
- Recall event number
- 98056