Open original FDA record ↗Recall record
- FDA identifier
- Z-0912-2026
- Company in FDA record
- Nipro Medical Corporation
- FDA classification date
- Dec 11, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Screw may loosen or fallout of Hemodialysis system
Source facts
- Action
- On October 27, 2025, Nipro issued a voluntary recall notification to affected consignees via E-Mail. In addition to provide information on the recall, Nipro asked consignees to take the following actions: 1. Please complete the enclosed SURDIAL DX Hemodialysis System Response Form to ensure the correct contact information for each machine to enable timely service appointments. 2. Pass this notice on to any facility that the Surdial DX Hemodialyis system has been transferred to. 3. Ensure all operators are aware of this recall notification. 4. Maintain a copy of the notification with affected machines. 5. Before each use, verify that the blood pump rotor assembly is securely fastened and operates smoothly without unusual vibration or looseness. 6. Monitor all impacted SURDIAL DX machines during operation for abnormal / unusual sounds, vibration or system error messages associated with the blood pump rotor assembly. 7. If any abnormality occurs, discontinue use immediately, safely remove the patient from the system following your facility protocols as appropriate.
- Initiated
- 2025-10-27
- Root cause
- Device Design
- Distribution
- US Nationwide distribution in the states of AL, AR, CA, FL, GA, IL, IA, MI, MN, MS, NJ, NY, NC, ND, OH, OR, PA, SC, SD, TX, WV, WC.
- Recall number
- Z-0912-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217152
- FDA classified
- 2025-12-11
- Recall event number
- 97914