Z-0912-2026: Hemodialysis System, Surdial DX REF: MC+SDX01 Rx Only

Recall record Z-0912-2026 for Nipro Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0912-2026
Company in FDA record
Nipro Medical Corporation
FDA classification date
Dec 11, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Screw may loosen or fallout of Hemodialysis system

Source facts

Action
On October 27, 2025, Nipro issued a voluntary recall notification to affected consignees via E-Mail. In addition to provide information on the recall, Nipro asked consignees to take the following actions: 1. Please complete the enclosed SURDIAL DX Hemodialysis System Response Form to ensure the correct contact information for each machine to enable timely service appointments. 2. Pass this notice on to any facility that the Surdial DX Hemodialyis system has been transferred to. 3. Ensure all operators are aware of this recall notification. 4. Maintain a copy of the notification with affected machines. 5. Before each use, verify that the blood pump rotor assembly is securely fastened and operates smoothly without unusual vibration or looseness. 6. Monitor all impacted SURDIAL DX machines during operation for abnormal / unusual sounds, vibration or system error messages associated with the blood pump rotor assembly. 7. If any abnormality occurs, discontinue use immediately, safely remove the patient from the system following your facility protocols as appropriate.
Initiated
2025-10-27
Root cause
Device Design
Distribution
US Nationwide distribution in the states of AL, AR, CA, FL, GA, IL, IA, MI, MN, MS, NJ, NY, NC, ND, OH, OR, PA, SC, SD, TX, WV, WC.
Recall number
Z-0912-2026
Recall status
Open, Classified
CDRH record ID
217152
FDA classified
2025-12-11
Recall event number
97914

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗