Open original FDA record ↗Recall record
- FDA identifier
- Z-0913-06
- Company in FDA record
- Bard Peripheral Vascular Inc
- FDA classification date
- May 31, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Manufacturing change that has introduced variability in the length of the inner stylet. When the inner stylet length is inserted into the cannula component, the exposed stylet length may be up to 8mm longer than expected.
Source facts
- Action
- Recall letters were sent to all consignees (Radiology Department Manager with a copy to the Hospital Administrator) via Federal Express overnight delivery with signature receipt required on May 1, 2006. Each letter included a Recall and Effectiveness Check Form for each consignee to fill out and fax to Bard Peripheral Vascular, Inc (BPV).
- Initiated
- 2006-05-01
- Root cause
- Other
- Terminated
- 2012-04-24
- Distribution
- The product was distributed to the following five states: 1) California, 2) Florida 3) New Jersey, 3) Missouri, 4) Texas, 5) Washington. The product was also distributed to Canada and Latin America. Note: units distributed to Latin America remained in Bard''s distribution system and have been returned.
- Recall number
- Z-0913-06
- Recall status
- Terminated
- CDRH record ID
- 45532
- FDA classified
- 2006-05-31
- Recall event number
- 35201