Skip to content

Public FDA record

Z-0913-06: Bard TruGuide Coxail Biopsy Needle with Depth Stop and additional Blunt Tip Stylet, Catalog Number C1816B

Recall record Z-0913-06 for Bard Peripheral Vascular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0913-06
Company in FDA record
Bard Peripheral Vascular Inc
FDA classification date
May 31, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Manufacturing change that has introduced variability in the length of the inner stylet. When the inner stylet length is inserted into the cannula component, the exposed stylet length may be up to 8mm longer than expected.

Source facts

Action
Recall letters were sent to all consignees (Radiology Department Manager with a copy to the Hospital Administrator) via Federal Express overnight delivery with signature receipt required on May 1, 2006. Each letter included a Recall and Effectiveness Check Form for each consignee to fill out and fax to Bard Peripheral Vascular, Inc (BPV).
Initiated
2006-05-01
Root cause
Other
Terminated
2012-04-24
Distribution
The product was distributed to the following five states: 1) California, 2) Florida 3) New Jersey, 3) Missouri, 4) Texas, 5) Washington. The product was also distributed to Canada and Latin America. Note: units distributed to Latin America remained in Bard''s distribution system and have been returned.
Recall number
Z-0913-06
Recall status
Terminated
CDRH record ID
45532
FDA classified
2006-05-31
Recall event number
35201

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗