Open original FDA record ↗Recall record
- FDA identifier
- Z-0913-2012
- Company in FDA record
- Stryker Communications Corp
- FDA classification date
- Jan 26, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Stryker has identified a potential structural issue with the Visum LED surgical light(s). During a complaint investigation, Stryker became aware that there was a possibility that incorrect assembly could result in the light head beginning to separate from the suspension arm. In the event the light separates and falls from the arm, it could lead to serious injury.
Source facts
- Action
- The firm sent "Urgent: Medical Device Correction" letters dated 11/07/2011 to their domestic and international customers. The letter informed customers of the remote risk of the light detaching from the arm and recommended continued use of the product. A representative will contact them about scheduling an appointment to inspect the lights. Contact Technical Services at 1-866-841-5663 for questions.
- Initiated
- 2011-11-07
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-08-17
- Distribution
- Worldwide Distribution - USA, and the countries of Australia, Canada, Hong Kong, Brazil, Sweden, India, Japan, Korea, Ecuador, Venezuela, Mexico, Denmark, Great Britain, and New Zealand.
- Recall number
- Z-0913-2012
- Recall status
- Terminated
- CDRH record ID
- 106527
- FDA classified
- 2012-01-26
- Recall event number
- 60812