Z-0913-2020: iChemVELOCITY, Urine Chemistry System, Catalog numbers: 700-7176-001, 700-7177-001, B75502 - Product Usa

Recall record Z-0913-2020 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0913-2020
Company in FDA record
Beckman Coulter Inc.
FDA record creation date
Jan 29, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Customers have reported incidents where two sets of results report the same Specimen Identifier (Specimen ID) with different results, different Medical Record Number (MRN) and different patient identification (demographics).

Source facts

Action
An Urgent Medical Device Recall letter was sent during the week of 10/28/2019 to the affected customers via mail and e-mail. The notice stated the misidentification issue as follows: Customers have reported incidents where two sets of results report the same Specimen Identifier (Specimen ID) with different results, different Medical Record Number (MRN) and different patient identification (demographics)." The following actions were provided to reduce the potential for this issue until a software update is available: - Follow good laboratory practice (GLP) and ensure that a unique identification system is in place. - Ensure that specimens are labeled with a unique Specimen ID and manual orders utilize a unique Specimen ID. - Do not reuse barcode number that identify different specimens. If barcode numbers must be reused, ensure results have been released from the LIS and the instrument Work List has been cleared. - Clear all manual Orders immediately after results are released, even if barcodes are used for subsequent runs (see iQ200 or iChemVELOCITY IFU, Chaper 7, Manual Orders).
Initiated
2019-10-28
Root cause
Software in the Use Environment
Terminated
2026-04-01
Distribution
Worldwide distributions.
Recall number
Z-0913-2020
Recall status
Terminated
CDRH record ID
178650
FDA record created
2020-01-29
Recall event number
84577

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗