Skip to content

Public FDA record

Z-0913-2022: Nextra Hammertoe Correction System 3.5 Middle and 3.2 Proximal Saw Cut SKU: NX-3532K-SC Orthopedic sur

Recall record Z-0913-2022 for Nextremity Solutions. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0913-2022
Company in FDA record
Nextremity Solutions
FDA record creation date
Apr 12, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery

Source facts

Action
Nextremity Solutions issued Urgent Medical Device Recall letter on 2/10/22 via mail and email to Distributors and Risk Mangers. Nextremity Solutions products with the referenced part and lot number should not be used and should be quarantined and returned to Zimmer Biomet for return to Nextremity Solutions. If you have affected product at your facility, locate and quarantine affected product in your inventory. Zimmer Biomet Sales representative will remove the affected product from your facility. Complete Medical Device Recall Return Response and send to CorporateQuality.PostMarket@zimmerbiomet.com, even if you do not have affected products at your facility. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Or email to CorporateQuality.PostMarket@zimmerbiomet.com.
Initiated
2022-02-10
Root cause
Nonconforming Material/Component
Terminated
2024-10-10
Distribution
Distribution US Nationwide, Japan, and the Netherlands
Recall number
Z-0913-2022
Recall status
Terminated
CDRH record ID
192384
FDA record created
2022-04-12
Recall event number
89764

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗