Skip to content

Public FDA record

Z-0914-06: EP-4 Computerized Stimulator -- 2 channel and 4 channel. The product consists of a touchscreen portable

Recall record Z-0914-06 for EP Medsystems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0914-06
Company in FDA record
EP Medsystems
FDA classification date
May 31, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Under certain circumstances the device may exhibit the following behaviors: 1)-Initiation of a rapid stimulation train without command, when the device is in use in conjuction with an electrosurgical unit or RF ablation unit-2)-Presence of a DC offset voltage on an output channel-3)-Inability to synchronize with a non-EP MedSystems recording system

Source facts

Action
The firm initiated the recall on April 4, 2006 through a notification letter sent to the electrophysiology lab managers at the healthcare institutions and two domestic distributors to which the device has been distributed. The firm used a traceable courier service to deliver notification of the recall.
Initiated
2006-04-04
Root cause
Other
Terminated
2007-04-24
Distribution
Product was shipped to hospitals/users throughout the US. The firm has a Veterans Affair contract number V797P-4650A. Also the firm has shipped the devices to Germany, United Kingdom, Netherlands, Italy, Japan, France, Saudi Arabia, Hong Kong, Syria, Malaysia, Sweden, Armenia, Australia, Canada, Iceland, S. Africa, and Taiwan.
Recall number
Z-0914-06
Recall status
Terminated
CDRH record ID
45507
FDA classified
2006-05-31
Recall event number
35206

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗