Open original FDA record ↗Recall record
- FDA identifier
- Z-0914-06
- Company in FDA record
- EP Medsystems
- FDA classification date
- May 31, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Under certain circumstances the device may exhibit the following behaviors: 1)-Initiation of a rapid stimulation train without command, when the device is in use in conjuction with an electrosurgical unit or RF ablation unit-2)-Presence of a DC offset voltage on an output channel-3)-Inability to synchronize with a non-EP MedSystems recording system
Source facts
- Action
- The firm initiated the recall on April 4, 2006 through a notification letter sent to the electrophysiology lab managers at the healthcare institutions and two domestic distributors to which the device has been distributed. The firm used a traceable courier service to deliver notification of the recall.
- Initiated
- 2006-04-04
- Root cause
- Other
- Terminated
- 2007-04-24
- Distribution
- Product was shipped to hospitals/users throughout the US. The firm has a Veterans Affair contract number V797P-4650A. Also the firm has shipped the devices to Germany, United Kingdom, Netherlands, Italy, Japan, France, Saudi Arabia, Hong Kong, Syria, Malaysia, Sweden, Armenia, Australia, Canada, Iceland, S. Africa, and Taiwan.
- Recall number
- Z-0914-06
- Recall status
- Terminated
- CDRH record ID
- 45507
- FDA classified
- 2006-05-31
- Recall event number
- 35206