Z-0914-2012: enGen(TM) Laboratory Automation Systems configured with Hettich Rotanta 46 RSC Robotic Centrifuges, Catal

Recall record Z-0914-2012 for Ortho-Clinical Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0914-2012
Company in FDA record
Ortho-Clinical Diagnostics
FDA classification date
Jan 26, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Ortho Clinical Diagnostics received complaints of centrifuge buckets disconnecting from their supports during centrifugation when using the Hettich Rotanta 46 RSC Robotic Centrifuges, Models 4815, 4816 and 4817, that may be configured with enGen(TM) Laboratory Automation Systems.

Source facts

Action
Ortho-Clinical Diagnostics sent an Urgent Product Correction Notification letter dated April 28, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to post the enclosed attachment near their centrifuge as a reference. Clean and carefully inspect the suspension groove area of the centrifuge buckets for cracks as part of their weekly maintenance procedure. Careful to check for micro-cracks that may not be easily detectable. If any cracks/micro-cracks, corrosion or damage are present, they must immediately discontinue using the centrifuge and contact OCD Customer Technical Services at 1-800-421-3311 to obtain replacements. OCD recommends that they do not exceed the manufacturer's relative centrifuge force (RCF) and run time limits for the blood collection devices used by their laboratory. Complete and return the attached Confirmation of Receipt Form. For questions regarding this recall call the Customer Service Center at 1-800-421-3311. .
Initiated
2011-04-28
Root cause
Component design/selection
Terminated
2018-07-02
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Australia, Canada, Chile, England, France, Germany, Italy, Mexico, Singapore and Spain.
Recall number
Z-0914-2012
Recall status
Terminated
CDRH record ID
106733
FDA classified
2012-01-26
Recall event number
60893

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗