Open original FDA record ↗Recall record
- FDA identifier
- Z-0915-2013
- Company in FDA record
- Steris Corporation
- FDA classification date
- Mar 6, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The control boards alarm set points and temperature tolerances allowed for erroneous activation of the AD (analog/digital) alarm as temperature fluctuated during operation. When this alarm occurs, the sterilizer locks up and cannot be used until serviced, causing Customer nuisance.
Source facts
- Action
- Steris Corporation initiated their recall on February 12, 2013. They will notify customers via onsite visits by STERIS Field Service Technicians to each affected Customer site. For further questions, please call (440) 392-7601.
- Initiated
- 2013-02-12
- Root cause
- Software design
- Terminated
- 2013-12-05
- Distribution
- Worldwide Distribution - USA including Canada and the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, SK, TN, TX, UT, VA, VT, WA, WI, and WV.
- Recall number
- Z-0915-2013
- Recall status
- Terminated
- CDRH record ID
- 116145
- FDA classified
- 2013-03-06
- Recall event number
- 64440