Z-0915-2020: ResMed Stellar, Non-invasive/invasive ventilators, Model: 100, 150

Recall record Z-0915-2020 for ResMed Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0915-2020
Company in FDA record
ResMed Ltd.
FDA record creation date
Feb 7, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Combination of software and a component failure may cause audible alarms not to operate properly, the alarm buzzer not work, for ventilators that have a failed electronic component and, that are stored without AC power connected for more than 36 hours leading to full depletion of the battery and, that powers on automatically when connected to AC power without pressing the power switch.

Source facts

Action
On 12/12/19, Urgent Field Safety Notifications were emailed to customers reinforcing user instructions. Use should be consistent with indications for use. Perform a functional test including alarms as described in the "Starting therapy" section in the User guide prior to each use. Press the power switch at the back of the device once to turn on the device. Check that the alarm sounds a test beep and the LEDs (visual indicator) for the alarm signal and the Alarm mute button flash. The firm is developing a device correction to address this issue. When this is available the firm will contact affected customers with specific actions at that time. Actions to be taken by healthcare professionals and users: The device is contraindicated in patients who are unable to endure more than brief interruptions in ventilation. It is important to assess changes in a patient's dependence on mechanical ventilation, particularly if their condition is progressive. If a patient is ventilator dependent, consider consulting their physician to discuss changing the patient to an appropriate device such as a life support device. Please continue to follow all patient and device information in the User and Clinical Guides, in particular, the following sections: Performing a functional test, Working with alarms, Internal battery, and Mobile use. As always, if there are any concerns regarding device behavior and operation, please contact your local customer service contact to organize a service. Actions to be taken by distributors and healthcare providers: Please forward this notice to your affected customers and physicians. For any questions, please call (855) 245-4640.
Initiated
2019-12-12
Root cause
Device Design
Distribution
U.S.: PA, MA, VA, TN, CT, IL, GA, FL, MT
Recall number
Z-0915-2020
Recall status
Completed
CDRH record ID
178585
FDA record created
2020-02-07
Recall event number
84551

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗