Open original FDA record ↗Recall record
- FDA identifier
- Z-0915-2023
- Company in FDA record
- Medtronic Neuromodulation
- FDA classification date
- Jan 6, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software anomalies, CP App messages: "Too Many Device Found", "Unexpected Device Error Code 1502", and "System Update Needed, Service Code 303"
Source facts
- Action
- The firm issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 12/03/2022 by mail. The notices explained the problems and the associated risks and provided the following: "Recommended Mitigation If the INS cannot be interrogated with Vanta PP App, contact Medtronic Technical Services at 1-800-707-0933. If needed, Medtronic will schedule an in-field service appointment with the managing SCS physician and the patient to diagnose and reset the INS." "Medtronic is working on a Vanta CP App software update that will resolve this issue and will communicate additional information when it becomes available."
- Initiated
- 2022-12-03
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution.
- Recall number
- Z-0915-2023
- Recall status
- Open, Classified
- CDRH record ID
- 197403
- FDA classified
- 2023-01-06
- Recall event number
- 91296