Reported results may have a positive bias of up to 113% because inadequate cleaning instructions may result in carry-over of the following reagents: Online TDM carbamazepine, Online TDM gentamycin, Online TDM phenytoin, Online DAT cocaine II, Online DAT methadone II, Online DAT opiates II and Online DAT THC II.
Source facts
Action
Foreign affiliates were notified via email letter dated 4/5/07 requesting that to notify their customers of the need to implement a special wash cycle to prevent reagent carry over.
Initiated
2007-04-05
Root cause
Other
Terminated
2008-04-30
Distribution
International Distribution ONLY-Australia, Canada and Germany.
Recall number
Z-0917-2007
Recall status
Terminated
CDRH record ID
51547
FDA classified
2007-06-09
Recall event number
37795