Skip to content

Public FDA record

Z-0917-2010: Sealapex Express (Part Number: 33639), root canal filling material. Manufactured by SybronEndo, Glendora

Recall record Z-0917-2010 for Sybron Dental Specialties. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0917-2010
Company in FDA record
Sybron Dental Specialties
FDA classification date
Mar 2, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Sybron Dental Specialties is initiating a recall on the Sealapex Xpress due to the presence of crystals in the catalyst. This crystallization may affect proper canal filling by preventing Gutta Percha Points from reaching the bottom of the root canal which may lead to an inadequate seal.

Source facts

Action
SybronEndo initiated an "Urgent Medical Device Recall" notification dated February 19 , 2009. Consignees were instructed to complete a "return form" and return any affected product in their inventory to the firm. Any product returned will be replaced at no charge. For further information, contact SybronEndo Customer Care at 1-800-346-3636.
Initiated
2009-02-23
Root cause
Process control
Terminated
2012-03-21
Distribution
Worldwide Distribution -- United States, Australia, Canada, United Arab Emirates, Belgium, Bulgaria, Germany, Spain, France, UK, Greece, Italy, Kuwait, Lebanon, Portugal, Russia, Sweden, Hong Kong, India, Malaysia and Taiwan.
Recall number
Z-0917-2010
Recall status
Terminated
CDRH record ID
85591
FDA classified
2010-03-02
Recall event number
53480

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗