Skip to content

Public FDA record

Z-0917-2012: VITROS Chemistry Products DT Calibrator Kit, REF/Catalog No. 195 7927, IVD --- Common/Usual Name: VITROS

Recall record Z-0917-2012 for Ortho-Clinical Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0917-2012
Company in FDA record
Ortho-Clinical Diagnostics
FDA classification date
Jan 27, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Ortho Clinical Diagnostics (OCD) was notified by customers of negatively biased quality control results obtained while using VITROS Chemistry Products NBIL DT Slides, GEN 72. OCD performed an internal investigation and confirmed that negatively biased NBIL results could occur for quality control, proficiency and patient samples. As a result, VITROS Chemistry Products DT Calibrator Kit Lots 37 &

Source facts

Action
Ortho Clinical Diagnostics (Johnson & Johnson Company) sent an Important Product Correction Notification letter dated October 13, 2011, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to recalibrate and instructing them to recalibrate VITROS Chemistry Products NBIL DT Slides, GEN 72 with updated DT calibrator values for the NBIL assay provided on CDM 0164. J&J Foreign affiliate consignees were notified by e-mail on October 13, 2011, informing them of the issue and of the availability of the updated CDM (0164). Consignees were instructed to complete and return the attached Confirmation of Receipt form upon receipt of the notification even if they do not use VITROS NBIL Slides in their facility. If the affected product was distributed outside the consignees facility they were instructed to forward the notification letter to affected customers. For any questions regarding this recall call 1-800-421-3311.
Initiated
2011-10-13
Root cause
Component design/selection
Terminated
2016-02-12
Distribution
Worldwide Distribution - USA (nationwide) Puerto Rico and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, and Venezuela
Recall number
Z-0917-2012
Recall status
Terminated
CDRH record ID
106492
FDA classified
2012-01-27
Recall event number
60799

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗