Z-0917-2020: Bard Conquest 40 PTA Dilation Catheter, 8mm X 20mm x 75cm (CQF7582), 8mm x 40mm x 75cm (CQF7584); Recomm

Recall record Z-0917-2020 for Bard Peripheral Vascular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0917-2020
Company in FDA record
Bard Peripheral Vascular Inc
FDA record creation date
Jan 30, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The catheter packaging incorrectly listed the balloon size as 8mm X 20mm instead of the intended size 8mm X 40mm. If undetected, the user will advance the longer than expected balloon to the lesion. The potential exists that upon inflation, non-diseased vessel, kidney or sections of the AV dialysis fistulae may be dilated, potentially causing a varying degree of injury. This may require additional medical or surgical intervention.

Source facts

Action
On 8/14/2019, a "Urgent Medical Device Recall Notification was sent to consignees via FedEx. Shortly after sending the original customer letter on 08/14/2019, the firm realized that customers may experience issues identifying the affected Lot Number in their inventory if they searched by the Product Code CQF7584. On 9/10/2019, the firm sent a 2nd notice explaining that although CQF7584 is the affected Product Code that they purchased, the actual Product Code listed on the package labeling is CQF7582. In addition to providing details about the recall products, the notification ask customer to do the following: 1. Our records show that your facility has purchased the product code affected by this voluntary recall. Do not use or further distribute any affected product. 2. Please check all inventory locations within your institution for the affected product code / lot number combination listed in the recall notice. If you have further distributed this product code / lot number, please immediately contact that location, advise them of the recall, forward these instructions and have them return the affected product to BPV. 3. Please remove any identified product from your shelves. 4. If you have used the affected product, complete and return the attached Recall and Effectiveness Check Form indicating no product will be returned. Once the product affected by this recall has been removed from your inventory: 5. Fill out the Recall and Effectiveness Check Form. Be sure to state the quantities and lot numbers of each recalled product that you intend to return. It is extremely important that we receive this information. 6. Email the completed Recall and Effectiveness Check Form to BDPI.CustomerSupportCenter@bd.com or fax it to BPV at 1-800-994-6772. 7. Once the Recall and Effectiveness Check Form has been completed and emailed to BDPI.CustomerSupportCenter@bd.com or faxed to BPV at 1-800-994-6772 and all information has been verified, the BPV C
Initiated
2019-08-14
Root cause
Labeling mix-ups
Terminated
2021-02-26
Distribution
US: NJ, NY, OH, PA, VA, SC, GA, AR, TX, MA, NC, TN, LA, AZ, MS, FL, AL,MD, CA, HI, MO, CO, WI, DC, KY, IL, MN, IN, KS, OK, CT, MT, WA, ND, IA, WA, WV, PR OUS: None
Recall number
Z-0917-2020
Recall status
Terminated
CDRH record ID
178949
FDA record created
2020-01-30
Recall event number
84659

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗