Skip to content

Public FDA record

Z-0917-2022: MM-TSM (Touch Screen Module/Secondary Controller/Bedside Controller), Part Numbers: 300002018101, 3000020

Recall record Z-0917-2022 for Volcano Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0917-2022
Company in FDA record
Volcano Corp
FDA record creation date
Apr 12, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Multi-Modality Touch Screen Modules (MM-TSM), used with integrated IVUS systems, may crack, or become damaged, which may affect their usability. They could become non-functional. Therefore, bottom protective beams will be replaced with full protective metal brackets.

Source facts

Action
On 02/03/22, correction notices were sent to customers who were asked to do the following: Notify all affected system users within the facility of this recall. The updated bracket improves protection on all sides of the MM-TSM touch screen that will prevent issues, cosmetic concerns, and maintain device integrity. If you are a distributor or have forwarded affected device(s) to another end user, it is imperative that all end-users with affected devices receive this Important Product Notice. Complete and return the customer response form. Customers with additional questions are encouraged to contact their local recalling firm representative or customer service: Phone: 1-800-228-4728, Option 2, Email: igtd.remotesupport@philips.com, Hours of Operation: Monday - Friday 8:00AM 5:00PM PST
Initiated
2022-02-03
Root cause
Device Design
Distribution
US: AZ, IL, NV, MI, FL, NC, AK, PA, WI, MN, CA, AR, NJ, TX, OR, NM, SD, CO, KY, LA, DE, MA, MT, MO, VA, AL, NY, WV, SC, IN, GA, IA, OH, UT, NH, MS, TN, OK, ID, MD, HI, RI, WY, DC, WA, KS, NE, ND, CT, VT. OUS: SK, IT, DK, GB, DE, SE, IN, KW, QA, PL, NL, CL, JP, BE, AU, FR, PF, PT, NZ, CA, ES, PK, CY, AT, IL, IE, CH, GR, SA, PS, LB, HK, IS, TH, AR, OM, AE, NO, ID, RO, RS, MM, ZA, SG
Recall number
Z-0917-2022
Recall status
Open, Classified
CDRH record ID
192244
FDA record created
2022-04-12
Recall event number
89698

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗