Open original FDA record ↗Recall record
- FDA identifier
- Z-0918-2014
- Company in FDA record
- Edwards Lifesciences, LLC
- FDA classification date
- Feb 3, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Edwards Lifesciences is recalling two lot numbers of the 26mm transcatheter heart valve crimper because they may not open sufficiently when the handle is rotated in the full open position. As a result, user may have difficulty fitting a 26mm Sapien transcatheter heart valve into the crimper orifice when preparing the valve for crimping.
Source facts
- Action
- Edwards Lifesciences issued an Urgent Product Notice to all customers. Letters were sent via Fed Ex on 01/20/14. The letter outlines steps to be taken to enable a valve to fit into the open orifice, and allow the valve to be crimped according to normal instructions. Customers were instructed to forward the Urgent Product Notification letter to necessary personnel within their organization. Customers were also instructed to acknowledge they have reviewed and understand the notification letter by signing and dating the form included with the letter. Customers with questions were instructed to contact their Edwards Clinical Specialist.
- Initiated
- 2014-01-20
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-03-24
- Distribution
- Nationwide Distribution
- Recall number
- Z-0918-2014
- Recall status
- Terminated
- CDRH record ID
- 125336
- FDA classified
- 2014-02-03
- Recall event number
- 67362