Recall classification
Class II

Reason stated in the source

There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.

Source facts

Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 12/6/24 was sent to customers. Actions to be taken by Customer/User: You can continue to use your Portrait" Mobile Monitoring Solution v1.1 product, since all potentially impacted systems were recently installed and the safety issue does not manifest until after 425 days of continuous use. Please follow the annual Maintenance Schedule described in Service Manual (PN 5929144) for your Portrait" Mobile to reboot the system. Rebooting the system by following this procedure resets the continuous days of use to zero. Please complete and return the attached Acknowledgement Form to FMI.36165@gehealthcare.com. After the Portrait Core Services v1.1 software is updated and corrected by GE HealthCare, please discontinue usage of any Portrait Core Services v1.1 software versions 1.1.0 or 1.1.1 and destroy any software media containing those versions. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2024-12-06
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution in the states of OH, TX and the countries of Australia, China, Czech Republic, France, Italy, Netherlands, Qatar, Singapore, UAE, UK.
Recall number
Z-0918-2025
Recall status
Open, Classified
CDRH record ID
211806
FDA classified
2025-01-15
Recall event number
96027

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗