Skip to content

Public FDA record

Z-0918-2026: Stryker MV3 bariatric bed, Part Number 5900000001

Recall record Z-0918-2026 for Stryker Medical Division of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0918-2026
Company in FDA record
Stryker Medical Division of Stryker Corporation
FDA classification date
Dec 11, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Stryker has issued a voluntary recall of certain MV3 beds and Arise 1000EX mattresses because some of these products have been sold in close proximity, suggesting a potential intent for them to be used together. These products have not been tested for compatibility and thus are not designed to be used together. The combination of both is not recommended by Stryker.

Source facts

Action
Stryker issued an URGENT: VOLUNTARY MEDICAL DEVICE SAFETY notice to its sales representatives on 11/13/2025 for hand delivery to its consignees on 11/14/2025. The notice explained the issue, potential hazard, and requested the following: "Actions needed: 1. Locate the MV3 and Arise 1000EX devices that have been shipped to your facility. Serial numbers are provided on the enclosed business reply form. 2. If an MV3 is being used together with an Arise 1000EX mattress, Stryker recommends the use to be discontinued upon receipt of this notification. 3. Complete the enclosed business reply form to confirm receipt of this notification. 4. Return your completed business reply form to Stryker by fax +1 269 488 8691 or email productfieldaction@stryker.com. 5. For those requesting a replacement of a compatible mattress, Stryker will contact you upon receipt of your Business Reply Form for next steps. 6. If you have loaned or sold any of the products listed in this letter, please forward a copy of this notice to the new users and advise us of their new location." For questions or concerns, contact Customer Service at +1 800 327 0770, Option 1 for Customer Support, Monday through Friday from 8:00 a.m. to 7:00 p.m. ET.
Initiated
2025-10-31
Root cause
Release of Material/Component prior to receiving test results
Distribution
US, nationwide
Recall number
Z-0918-2026
Recall status
Open, Classified
CDRH record ID
217312
FDA classified
2025-12-11
Recall event number
97953

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗