Open original FDA record ↗Recall record
- FDA identifier
- Z-0919-2011
- Company in FDA record
- Xsensor Technology Corporation
- FDA classification date
- Jan 18, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The plastic enclosures for the X3 Display and X3 PRO Pressure Mapping System has a UL94 HB flammability rating while the medical equipment specification UL 60601-1 requires V-2 rated plastic for electronic enclosures.
Source facts
- Action
- On July 6, 2009, Xsensor sent an "URGENT: MEDICAL DEVICE RECALL" notifications to their consignees/customers. The consignees/customers were informed that the incorrect material of the plastic enclosures was used. The incorrect plastic had a UL94 HB flammability rating while the medical equipment specification UL 60601-1 required V-2 rated plastic for electronic enclosures. The firm provided prepaid shipping box to consignees/customers with instructions for returning the components. The components were refurbished and returned to consignees for free of charge. For further information please contact XSENSORs Customer Support at 1-866-927-5222.
- Initiated
- 2009-07-20
- Root cause
- Component change control
- Terminated
- 2011-01-31
- Distribution
- Devices were distributed to one distributor, medical facilities, and rehab centers through out the US, including following states: AL, AR, AZ, CA, CO, FL, GA, IA, IL, KS, ME, MI, MN, MO, NC, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, and WI.
- Recall number
- Z-0919-2011
- Recall status
- Terminated
- CDRH record ID
- 87910
- FDA classified
- 2011-01-18
- Recall event number
- 54229