Reason stated in the source
The Sirus Drill and Flexible Shaft instruments may not be adequately cleaned when utilizing the standard cleaning instructions potentially resulting in infection and subsequent complications.
Source facts
- Action
- Review the notification and ensure affected personnel are aware of contents. Assist Zimmer Biomet sales representative in locating and quarantining affected product. Sales representative will remove product from facility. Complete certificate of acknowledgement and return to Zimmer Biomet. Distributors are asked to locate and quarantine all affected product in inventory and from affected hospitals in their territory.
- Initiated
- 2017-10-03
- Root cause
- Labeling design
- Terminated
- 2018-08-20
- Distribution
- Distributed in 31 states: AL, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, KY, MA, MD, MI, MN, MO, NC, NJ, NM, NY, OH, OR, PA, TN, TX, VA, VT, WA, WI and the District of Columbia.
- Recall number
- Z-0919-2018
- Recall status
- Terminated
- CDRH record ID
- 161655
- FDA record created
- 2018-03-06
- Recall event number
- 79206