Skip to content

Public FDA record

Z-0919-2021: Brain Heart Infusion Agar with 6 ug/ml Vancomycin Cat no: G14 Lot no: 472698 Manufactured: 2020-10-29 Exp

Recall record Z-0919-2021 for Hardy Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0919-2021
Company in FDA record
Hardy Diagnostics
FDA record creation date
Jan 25, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Complaints were received reporting of Staphylococcus aureus (ATCC 29213) breakthrough when the strain should have been inhibited.

Source facts

Action
On 12/21/2020, the firm sent an "URGENT: Medical Device Recall Notification" letter to customers via first class mail, email and/or facsimile informing them that seven customer complaints received of breakthrough of negative controls. Customers are instructed to: 1. Discard any remaining stock. The product should be destroyed according to local regulation as non-hazardous waste if un-inoculated, or as medical waste if inoculated. 2. Retain the Recall Notification and to notify any and all employees who may use the product. If the product was further distributed, to notify the consignees the product was distributed to. 3. Fill out the Product Recall Customer Inventory Form and to return the form within five (5) business days via e-mail to JohnsonL@HardyDiagnostics.com or by fax to (805)266-2222, option 2 or (805)361-5050 or via e-mail at TechnicalServices@HaryDiagnostics.com. For any questions or further information contact Technical Services Department at 800-266-2222 ext. 5598
Initiated
2020-12-21
Root cause
Under Investigation by firm
Terminated
2022-09-28
Distribution
U.S.: CA, UT, MI, Puerto Rico, IN, CT, AZ, GA, MA, TN, WA O.U.S. None
Recall number
Z-0919-2021
Recall status
Terminated
CDRH record ID
185320
FDA record created
2021-01-25
Recall event number
87091

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗