The diodes in the control panel may fail, which could result in the thermal breakdown of the circuit board inside of the control box, and cause the bed to cease functioning.
Source facts
Action
Consignees were sent a fax on 5/17/05 advising them of the problem and instructing them to lock-out the HILO pendant function on the attendant control pad until the firm visits their facility and corrects their beds.
Initiated
2005-05-17
Root cause
Other
Terminated
2005-11-17
Distribution
Nationwide.
Recall number
Z-0920-05
Recall status
Terminated
CDRH record ID
39115
FDA classified
2005-06-22
Recall event number
32093