Probe Wash: Reported results may have a positive bias because inadequate analyzer probe wash instructions may result in carry-over from the following assays: Creatinine Kinase L and Creatine Kinase-MBL.
Source facts
Action
Consignees were issued a recall letter dated 4/30/07 instructing them to perform a R2 reagent probe wash when using CKL and CKMBL with the vancomycin or valproic acid assays.
Initiated
2007-04-30
Root cause
Other
Terminated
2007-12-17
Distribution
Nationwide.
Recall number
Z-0920-2007
Recall status
Terminated
CDRH record ID
51551
FDA classified
2007-06-09
Recall event number
37796