Open original FDA record ↗Recall record
- FDA identifier
- Z-0920-2017
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Jun 21, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.
Source facts
- Action
- Siemens mailed on December 01, 2016 a Customer Safety Notice to affected customers to inform them of a corrective action that is being performed to prevent a possible hazard to persons and equipment. The notice explained the underlying issue, the effect the issue has on the system, the associated risks and actions they will take to correct the issue. In addition, customers will receive a Customer Safety Notice via Hand Delivery from the Siemens Customer Service Engineer at the scheduled appointment for the corrective installation.
- Initiated
- 2016-11-22
- Root cause
- Nonconforming Material/Component
- Terminated
- 2017-05-30
- Distribution
- Distributed to: IL IL GA SC SD OK NY MI IN MD AR
- Recall number
- Z-0920-2017
- Recall status
- Terminated
- CDRH record ID
- 151464
- FDA record created
- 2017-06-21
- Recall event number
- 75860