Skip to content

Public FDA record

Z-0920-2021: SERF Novae Dual Mobility Cup Insert 51/28- ApeX-LNK Poly Liner. Intended for use with the Apex Modular Ac

Recall record Z-0920-2021 for OMNIlife science Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0920-2021
Company in FDA record
OMNIlife science Inc.
FDA record creation date
Jan 26, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Acetabular Liners have a higher than acceptable oxidation limit, potential for patients to experience a higher wear rate for acetabular liners and require a revision surgery

Source facts

Action
Omni Life Initial communication was sent by email via a quarantine form on 12/09/2020. On 12/22/20 a notification packet detailing parameters of recall sent via email and overnight mail. Acknowledgement of receipt is requested. Actions to be taken by the Customer/User " Immediately quarantine lots 28413, 28624 and 28572 " Return Product to OMNI at 480 Paramount Drive, Raynham MA 02767 " This advisory notice only includes the lots identified above. It does not affect other lots. You must acknowledge this notice by returning the attached Acknowledgement and Receipt Form by FAX to 508-819-3390 or my email to Omni-cs@coringroup.com.
Initiated
2020-12-21
Root cause
Under Investigation by firm
Terminated
2021-06-10
Distribution
AL, ID, MD, NY, OK
Recall number
Z-0920-2021
Recall status
Terminated
CDRH record ID
185144
FDA record created
2021-01-26
Recall event number
87062

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗