Open original FDA record ↗Recall record
- FDA identifier
- Z-0920-2025
- Company in FDA record
- SIN SISTEMA DE IMPLANTE NACIONAL S.A
- FDA classification date
- Jan 15, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Dental implant box incorrectly labels the diameter as 4,5 X 18,0 mm, but the internal label correctly labels the diameter as 4,0 X 18,0 mm; if the dentist drills the bone with the larger diameter stated on the box, the smaller diameter implant, will become loose and will not fit, which may necessitate an additional surgery.
Source facts
- Action
- On 9/1/2024, "URGENT: MEDICAL DEVICE RECALL" notices were emailed to customers and distributors who were told the following: The product should not be used. It does not pose a handling risk and should be separated and discarded. Complete and return the Medical Device Recall Return Response Acknowledgement and Receipt Form via email to: brian@sindentalusa.com. Customer with questions may contact the company via telephone at 0800-770-8290 between the hours of 9 and 19 (Brasilia time) or email the firm at anna.lopes@sinimplantsystem.com.
- Initiated
- 2024-09-01
- Root cause
- Process control
- Distribution
- US distribution to states of: KY, CO, CA, NY, ID, NJ, FL, TX, KY, IL, OH; AND OUS (foreign) country of Brazil.
- Recall number
- Z-0920-2025
- Recall status
- Open, Classified
- CDRH record ID
- 211702
- FDA classified
- 2025-01-15
- Recall event number
- 95983