Skip to content

Public FDA record

Z-0920-2026: Hardware configuration of the BeneVision Central Monitoring System (CMS), marketed as BeneVision Distribu

Recall record Z-0920-2026 for Mindray DS USA, Inc. dba Mindray North America. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0920-2026
Company in FDA record
Mindray DS USA, Inc. dba Mindray North America
FDA classification date
Dec 12, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When the Worstation of the BeneVision DMS has a specific hardware configuration, the computer may experience audio playback failure or screen freezing.

Source facts

Action
Mindray notified consignees of the recall on about 11/18/2025 via certified mail. Consignees were informed that they may continue to use the Central Monitoring System. If the HP Pro Mini 400 G9 Desktop PC used by the WS experiences the audio failure or screen freezes, restart the HP Pro Mini 400 G9 Desktop PC. Consignees were instructed to ensure applicable clinical staff are aware of the issue and letter and shared with any customer if further distributed. Consignees were also asked to complete and return the Acknowledgement and Receipt form provided. Mindray will provide a patch update for the audio driver of the affected HP Pro Mini 400 G9 Desktop PCs located at each facility and is anticipated to take approximately 10-15 minutes per HP Pro Mini 400 G9 Desktop PC. A representative from Mindray's Care team will contact consignees to schedule the update.
Initiated
2025-11-18
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Recall number
Z-0920-2026
Recall status
Open, Classified
CDRH record ID
217345
FDA classified
2025-12-12
Recall event number
97969

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗