Skip to content

Public FDA record

Z-0921-2013: 3.75mm x 12mm Fixed Cervical Screws. Components of the Genesys Spine Anterior Cervical Plate System Pro

Recall record Z-0921-2013 for Genesys Orthopedic Systems, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0921-2013
Company in FDA record
Genesys Orthopedic Systems, LLC
FDA classification date
Mar 8, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Manufacturing residue may be present on the cervical screws. Use of this product may result in patient infection, delayed fusion or non-fusion (pseudarthrosis), inflammatory response, or revision surgery.

Source facts

Action
Genesys Orthopedic Systems, LLC sent a Recall Notification letter dated February 20, 2013 to all affected customers. The letter identified the affected product, problem and necessary actions to be taken. Customers were instructed to immediately return the affected product to the Genesys Spine office and complete and return the enclosed Response Form as soon as possible. For questions call 512-381-7070.
Initiated
2013-02-20
Root cause
Employee error
Terminated
2013-09-09
Distribution
USA Nationwide Distribution including the states of: CA and TX.
Recall number
Z-0921-2013
Recall status
Terminated
CDRH record ID
116108
FDA classified
2013-03-08
Recall event number
64421

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗