Open original FDA record ↗Recall record
- FDA identifier
- Z-0922-2020
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA record creation date
- Jan 30, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Cobasinfinity laboratory solution Version 2.4.1 through Version 2.5.4 Using a CommServer Driver - Incorrect Alarm Mapping
Source facts
- Action
- Roche issued Urgent Medical Device Correction TP-00791 mailed via UPS Ground (receipt required) on 10/8/2019. The letter identifies the problem, health risk and action to take: Complete the attached fax form (TP-00792) and fax or email it according to the instructions on the form. " File this Urgent Medical Device Correction (UMDC) for future reference Roche will resolve this issue in a future software update. In the next couple of weeks, a Roche Representative will be contacting you to adjust alarm code mapping as needed per your specific configuration. Contact the Roche Support Network Customer Support Center 24 hours a day, seven days a week at 1-800-526-2272 if you have questions. The UMDC will also post to the diagnostics.roche.com website
- Initiated
- 2019-10-08
- Root cause
- Software design
- Terminated
- 2020-10-15
- Distribution
- US Nationwide in the states of AL, CA, GA, IA, IL, KY, MI, MO, MT, NJ, NY, OR, TN
- Recall number
- Z-0922-2020
- Recall status
- Terminated
- CDRH record ID
- 177289
- FDA record created
- 2020-01-30
- Recall event number
- 84128