A potential issue with ARTIS One systems was identified. In very
rare situations, the first x-ray release following patient registration may be performed with incorrect
copper prefiltration. If the described issue occurs, the applied radiation dose-rate may be higher
than intended by the user.
Source facts
Initiated
2025-01-02
Root cause
Radiation Control for Health and Safety Act
Distribution
23 Distributed in the US
Recall number
Z-0922-2025
Recall status
Open, Classified
CDRH record ID
211984
FDA classified
2025-01-29
Recall event number
96088