Z-0923-2011: K-ASSAY Alpha-1 Microglobulin Reagent Kit. Kit contains: 1 vial Reagent 1, Buffer 50 mL, and 1 vial reag

Recall record Z-0923-2011 for Kamiya Biomedical Company, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0923-2011
Company in FDA record
Kamiya Biomedical Company, LLC
FDA classification date
Jan 19, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Standardization of the reagent performance data did not match the standardization used to assign the calibrator values.

Source facts

Action
Kamiya Biomedical Company, LLC sent an urgent medical Device Recall letter dated September 17, 2008, and the corrected package insert via e-mail. . The letter identified the product, the problem, and the action the customer should take. Customers were instructed to replace their previous package inserts with the corrected version and asked to sign and FAX the letter back once that had been completed. As one customer is also a distributor, the recalling firm told the this distributor's representative to advise their single customer of this correction and replace the package insert, and return the signed letter back to the firm. On September 22, 2008, both customers were contacted by phone and reminded to fax the signed letters once completed. For questions regarding this recall call 1-800-222-0342.
Initiated
2008-09-17
Root cause
Employee error
Terminated
2011-01-20
Distribution
Kits were distributed to one laboratory in MN and to one distributor in PA who further distributed product to one customer.
Recall number
Z-0923-2011
Recall status
Terminated
CDRH record ID
74651
FDA classified
2011-01-19
Recall event number
50074

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗