Skip to content

Public FDA record

Z-0923-2025: 8MM, Tenaculum Forceps REF 470207 Instrument used with the da Vinci X/Xi Surgical Systems in a wide vari

Recall record Z-0923-2025 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0923-2025
Company in FDA record
Intuitive Surgical, Inc.
FDA classification date
Jan 16, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to an increase in complaints for pitch cable failures related to forceps and small Graptor instruments

Source facts

Action
On 12/19/2024, the firm sent an "Field Safety Notice Urgent Medical Device Notification" via FedEx (US), email/hand-delivery (OUS), informing customers that Intuitive Surgical has observed an increase in complaints regarding pitch cable failure on the Tenaculum Forceps and Small Graptor retractor devices. A failure of the pitch cable can be fraying or complete break. Customers are instructed: 1. As a reminder, when using the Tenaculum Forceps and the Small Graptor instruments, refer to and follow the instructions, warnings and cautions provided in the General Overview and EndoWrist Instrument chapters of the da Vinci X/Xi Instruments and Accessories User Manual and Reprocessing Instructions User Manual. 2. Display this communication with affected systems; ensures it is in a place likely to be seen/viewed by operators. 3. Complete the Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form. 4. Retain a copy of this letter and the acknowledgement form for your files. 5. Inform Intuitive of any Adverse Events*/Serious Incidents** or quality problems concerning the use of the subject devices via the standard complaint process. For questions/information/further support - contact Clinical Sales Representative or Customer Service: "North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: customerservice@intusurg.com. "Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or eucs@intusurg.com "South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or support.korea@intusurg.com "Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or csjapan@intusurg.com "India: +1-800-103-6952 (9 am to 6 pm IT) "Taiwan: +0800-86-8181 (9 am to 6 pm CT)
Initiated
2024-12-19
Root cause
Device Design
Distribution
U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WV, WI, and WY. The countries of Canada, Austria, Belgium, Denmark, Finland, France, French Guiana, Germany, Guadeloupe, Ireland, Luxembourg, Martinique, Monaco, Netherlands, Norway, Reunion, Sweden, Switzerland, United Kingdom, India, Japan, South Korea, Taiwan, Brazil, Argentina, Australia, Bulgaria, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Ecuador, Greece, Hong Kong, Hungary, Iceland, Israel, Italy, Kuwait, Lebanon, Malaysia, Malta, Mexico, New Zealand, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Thailand, Turkey, United Arab Emirates, Venezuela, Vietnam.
Recall number
Z-0923-2025
Recall status
Open, Classified
CDRH record ID
211572
FDA classified
2025-01-16
Recall event number
95939

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗