It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro
exams. This can happen in case the framerate is increased on the NX workstation when ABS is not
enabled.
Source facts
Action
AGFA NV notified Customers via a letter titled "IMPORTANT PRODUCT INFORMATION LETTER / DR 800", included Product Name, Symptom, Cause, Actions to be taken customer and manufacturers respectively: "Always enable ABS for pulsed fluoro exams.
Actions being taken by the manufacturer:
Agfa/local business partner will contact you to install a new generator software."
Initiated
2025-11-18
Root cause
Radiation Control for Health and Safety Act
Distribution
U.S. Nationwide distribution.
Recall number
Z-0923-2026
Recall status
Open, Classified
CDRH record ID
217616
FDA classified
2025-12-22
Recall event number
98101