Open original FDA record ↗Recall record
- FDA identifier
- Z-0924-2011
- Company in FDA record
- Mizuho Orthopedic Systems Inc
- FDA classification date
- Jan 20, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Device component spar gear box locks up and may result in failure of push pull tube, which may result in release of breaking and impact.
Source facts
- Action
- All customers will be contacted to arrange for removal and shipping arrangements. Verification will be done by the Repair and Verification instructions for the leg spars, via registered letters. *******UPDATED INFO---The firm sent another notification letter---dated 12/22/10---which states in part: "****The upgrade involves the correction of your 6850 PROfx spars (6850-2100) at your facility by a Qualified Technician. The spars shall be upgraded to remove the 6850-2123, Drive Bracket, Spar Push Rod and replace it with the 6850-2170, Drive Bracket, Spar Push Rod. Upon completion of the upgrade, the 6850-2123, Drive Bracket, Spar Push rod shall be returned to Mizuho OSI for processing.*******
- Initiated
- 2010-03-30
- Root cause
- Device Design
- Terminated
- 2011-05-17
- Distribution
- Worldwide Distribution -- USA & Germany.
- Recall number
- Z-0924-2011
- Recall status
- Terminated
- CDRH record ID
- 94677
- FDA classified
- 2011-01-20
- Recall event number
- 56859