Skip to content

Public FDA record

Z-0925-2007: LIFEPAK 20 defibrillator/monitor. The device is an AC powered defibrillator device with battery backup. T

Recall record Z-0925-2007 for Medtronic Emergency Response Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0925-2007
Company in FDA record
Medtronic Emergency Response Systems, Inc.
FDA classification date
Jun 9, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

White Screen Lock up: LifePak 20 defibrillators with software version 48 or version 52 may experience an intermittent timing issue during the power on sequence. This may lock up the device resulting in the monitor screen going white, and may cause a potential delay or prevention of shock treatment.

Source facts

Action
On 04/27/2007 firm sent letters via return receipt to all domestic and foreign customers. The letter recommends for customers to turn the defibrillator power off and back on to restore normal operation in the event of a white screen condition. In addition, the letter advises that a representative from the firm will contact customer to make arrangements to provide a software upgrade to correct the issue.
Initiated
2007-04-27
Root cause
Other
Terminated
2012-03-23
Distribution
Distribution is Worldwide. Devices are used by healthcare providers in hospital or healthcare clinic settings.
Recall number
Z-0925-2007
Recall status
Terminated
CDRH record ID
51986
FDA classified
2007-06-09
Recall event number
37864

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗