Open original FDA record ↗Recall record
- FDA identifier
- Z-0925-2013
- Company in FDA record
- Biomet, Inc.
- FDA classification date
- Mar 8, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup during impaction. If alternative inserters are not available this could cause a surgical delay greater than 30 minutes.
Source facts
- Action
- BIOMET sent an Urgent Recall Notice beginning February 13, 2013, to all affected customers. The notice included possible adverse events and directions to locate and remove identified devices by returning a Fax Back Response Form to 574-372-1683 and obtaining an RGA # prior to return of product to Biomet, Inc. 56 East Bell Drive Warsaw, IN 46580. Questions should be directed to 574 372-1570 M-F, 8 a.m.-5.p.m. For questions regarding this recall call 574-267-6639, ext 1676.
- Initiated
- 2013-02-05
- Root cause
- Process control
- Terminated
- 2013-09-30
- Distribution
- Worldwide Distribution - No USA distribution -Internationally to Canada, Japan, and Chile
- Recall number
- Z-0925-2013
- Recall status
- Terminated
- CDRH record ID
- 115921
- FDA classified
- 2013-03-08
- Recall event number
- 64354